Product Description
REVO-2 head kit
Includes: REVO-2®, RSP2, RSH250, 45 mm diameter datum ball
The 5-axis REVO-2 system incorporating the REVO-2 head provides fast, ultra-high accuracy positioning.
REVO-2 system overview
The system comprises the following elements:
- REVO-2 head
- RSP2 2D tip sensing probe and associated stylus holders and accessories
- RSP3 3D probe and associated accessories
- UCC universal CMM controller
- REVO-2 PCI interface card (for UCC)
- SPA servo power amplifier
- Air filter unit
REVO-2 - 'tip sensing' probe technology
- Enclosed laser directed onto a reflector at the stylus tip.
- The stylus touches the part and bends.
- The reflector is displaced.
- The altered return path of the laser is sensed by a PSD.
- The exact tip position is known because the reflector and the stylus ball are close together.
- Stylus wear is minimized by using a low scanning force.
REVO-2 features and benefits
- Incorporates Renscan5™ 5-axis scanning technology minimizing CMM motion and the associated CMM dynamic errors
- Increased measuring speed, up to 500 mm/s resulting in increased measurement throughput
- Data collection rates up to 6,000 points per second
- Infinite positioning 5-axis motion reduces non-productive transitions between features
- Rapid calibration with all positions inferred means more time measuring
- Maximum reach up to 500 mm with maintained effective working length
- Standard M2 styli for convenience
- Probe and stylus changing capability allowing flexibility and future probing technology capability
Hardware integration
- The UCC is fundamental to the REVO-2 system
- The UCC controller features Renscan5™ scanning routines particular to 5-axis motion and scanning
- SPA is a servo power amplifier used to drive the head and CMM
- MCU is the multi-function hand control unit required for the system
Software integration
- The Renishaw UCCserver™ software application will provide the interface for REVO control
- The UCCserver™ is based on the I++ DME command protocol
Specification
Operating temperature | +10 °C to +40 °C (+50 °F to +104 °F) |
---|---|
Storage temperature | -25 °C to +70 °C (-13 °F to +158 °F) |
Weight (excluding probe and cables) | 2.1 kg (4 lb 10 oz) |
Height (overall) | 176.5 mm (6.95 in). Quill to probe mount, including 4 mm mounting plate |
B-axis | 92 mm × 93.5 mm (3.62 in × 3.68 in) |
A-axis swept diameter | 116 mm (4.56 in) |
Movement speed | 3 revs/sec |
A-axis rotation angle | -5° to +120° (for measurement with RSP2 and RVP) -100° to +120° (used for calibration and measurement with RSP3 and SFP1) |
B-axis rotation angle | Continuous |
Angular resolution | 0.02 arc sec (0.01 µm / 100 mm) |
Bearings | Air - consumption rate up to 35 l/min (5 bar to 8.5 bar) (Refer to user's guide for detail) |
Change rack system | RCP TC-# for probe changing / RCP for stylus changing; VPCP and VMCP for vision system |
Mounting | Quill mounting in the vertical orientation only (adaptor may be required) |